| The digital
measuring principle |
| The pendulum, suspended by the Archimedes
helical spring, is mounted between two electrodes. Depending on the inclined position of
the system, the pendulum will swing out of the zero-position and by that, changing the
capacity between the pendulum and the two electrodes. These capacities will be transformed
into different frequencies through the RC-oscillator. The ratio of the two frequencies
available will be used as the primary signal for detecting the required angle. (The system
is patent protected in most countries) |
| WK-34-5118 CLINO 2000 |
 |
|
- A highest possible precision over the large measuring range
of +/- 45° with integrated temperature compensation.
- Effortless zero adjustment by using the integrated software
and a reversal measurement.
- Easy to calibrate due to
implemented software guidance and the calibration aids as part of the delivery.
- Large digital display with the
advantage to set all the measuring units commonly used.
- Built-in possibility to connect an
additional instrument for differential measurement or ZEROTRONIC sensors by using
the serial port.
- Rugged stainless steel housing
with prismatic bases.
- Built-in cross vial for easy alignment of the secondary
vertical setting direction in order to eliminate twist error".
- State of the Art digital technology.
- The instrument is fully compatible with the whole range of
digital sensors of WYLER AG.
- Powered by common 1.5 V batteries,
rechargeable batteries or with mains adapter.
- Fulfills the strict CE requirements (Immunity against
electromagnetic smog).
- As an option magnetic inserts are available.
|
Features:
Suitable for mobile use; compact sturdy stand alone unit with possibility to
adjust all commonly used measuring units; not sensitive to magnetic field. |
Specifications
CLINO 2000 |
Measuring range |
±45
Degrees |
Options |
±10 Deg.
/ ±30 Deg. / ±60 Deg. |
Calibration |
21
Setting points with temperature compensation |
Settling time
(DIN 2276) |
< 5
Seconds |
Resolution
(DIN 2276) |
5 Arcsec
(0.025mm/m) |
Limits of error
(DIN 2276) |
< 5
Arcsec + 0,07% R.O. |
Limits of error after quick calibration
(DIN 2276) |
< 30
Arcsec |
Data connection |
RS 485,
asynchron, 7 Bit, 2 Stop bits, no parity, 9600 Baud |
Power supply with batteries (Lifetime) |
2 x Size
AA 1.5V Alkaline (40-60 Std) or |
External power supply |
-12 ...
+48 V DC / 200-500mW |
Housing (Weight) |
Stainless
steel
150 x 150 x 35 mm (3 kg) |
Temperature range |
0 to
40°C |
Storage temperature |
-20 to
+70°C |
|
PRICE